Compatible Disposable Transducer Invasive DPT Blood IBP for PVB

Product Details
Customization: Available
Type: Surgical Supplies Materials
Material: High Polymer Material and Electronic Chips
Still deciding? Get samples of US$ 6/Piece
Request Sample
Manufacturer/Factory & Trading Company
Gold Member Since 2021

Suppliers with verified business licenses

Audited Supplier

Audited by an independent third-party inspection agency

Number of Employees
160
Year of Establishment
2009-03-20
  • Compatible Disposable Transducer Invasive DPT Blood IBP for PVB
  • Compatible Disposable Transducer Invasive DPT Blood IBP for PVB
  • Compatible Disposable Transducer Invasive DPT Blood IBP for PVB
  • Compatible Disposable Transducer Invasive DPT Blood IBP for PVB
  • Compatible Disposable Transducer Invasive DPT Blood IBP for PVB
  • Compatible Disposable Transducer Invasive DPT Blood IBP for PVB
Find Similar Products
  • Overview
  • Product Description
  • Detailed Photos
  • Company Profile
  • Exhibition
  • Certifications
  • Packaging & Shipping
Overview

Basic Info.

Model NO.
JIBPT-01-PVB
Ethylene Oxide Sterilization
Ethylene Oxide Sterilization
Quality Guarantee Period
Three Years
Group
All
Logo Printing
With Logo Printing
Certificate
CE/ISO13485
Properties
Medical Materials & Accessories
Transport Package
Blister Box
Specification
270*175*54mm
Trademark
JCR
Origin
China
HS Code
9018909991
Production Capacity
400000/Month

Product Description

Product Description

Properties
Medical Materials & Accessories
Product Name
   IBP Transducers for Edward Type
Brand Name
JCR
Model Number
JIBPT-01-PVB

Detailed Photos

Main technical parameters and working environment requirements:
Disposable Pressure Transducer
Barometric pressure
70~106Kpa
Relative humidity
10~90% (non-condensing)
Operating pressure range
-50 ~ + 300 mmHg
Sensitivity
5.0μV/V/mmHg ±3%
Nonlinearity and hysteresis
full-scale reading±1.5%
Input impedance
1200Ω ~3200Ω
Zero pressure offset
-20mmHg~+20mmHg
Thermal offset shift
≤±0.3mmHg/°C
Offset drift
After warm-up for 20 seconds, drift within 2 mmHg within 8 hours
Thermal span shift
≤±0.1%/°C
Frequency response
Standard pressure set (48 "/ 12") is 40 Hz; Separate sensor > 200 Hz
Defibrillator withstand
Defibrillation depends on the final connection of the equipment.
Leakage current
Leakage current is dependent on the final connection of the equipment
Overpressure load
-400~+4000mmHg
Shock resistance
Withstand falling from one meter for three times
 
Light sensitivity
When exposed to a 3400°K tungsten light, candle light at 3000 feet, it's less than 1 mmHg under rated voltage.
Storage and transport environment requirements
- 20 ° C~ + 60 ° C, < 90%(non-condensing), no corrosive gas and indoor ventilation, and to avoid high temperature and cold.
Contact with the human body
≤168h
Intra-system velocity
Flow
2-5mL/h
Intra-system velocity refers to the velocity under 6.00 VDC and 25°C, unless there are additional instructions.
Product shelf-life
3 years
Compatible Disposable Transducer Invasive DPT Blood IBP for PVBCompatible Disposable Transducer Invasive DPT Blood IBP for PVBCompatible Disposable Transducer Invasive DPT Blood IBP for PVBCompatible Disposable Transducer Invasive DPT Blood IBP for PVB
Compatible Disposable Transducer Invasive DPT Blood IBP for PVB
Compatible Disposable Transducer Invasive DPT Blood IBP for PVB

Company Profile

Compatible Disposable Transducer Invasive DPT Blood IBP for PVB

 

Exhibition

 

Compatible Disposable Transducer Invasive DPT Blood IBP for PVB

Certifications

Compatible Disposable Transducer Invasive DPT Blood IBP for PVB

Packaging & Shipping

Compatible Disposable Transducer Invasive DPT Blood IBP for PVBCompatible Disposable Transducer Invasive DPT Blood IBP for PVB
Compatible Disposable Transducer Invasive DPT Blood IBP for PVB

Send your message to this supplier

*From:
*To:
*Message:

Enter between 20 to 4,000 characters.

This is not what you are looking for? Post a Sourcing Request Now
Contact Supplier