Are you in need of a reliable and accurate solution for measuring arterial and venous blood pressure in ICU Anesthesiology? Look no further than our Disposable Pressure Transducer Kit! Designed to provide consistent and precise readings, this kit is a must-have for medical professionals.
Our Disposable Pressure Transducer Kit comes with a range of features to meet your specific needs. With options for flushing at 3cc/hr, 30cc/hr, or without flush valves, you can customize the kit to suit your requirements. It includes essential components such as infusion sets, pressure monitoring lines, and stopcocks, ensuring you have everything you need for seamless blood pressure measurement.
At Shenzhen JCR Medical Technology Limited Company, we take pride in delivering high-quality products. Our Disposable IBP Transducer Kit with Logo Printing is no exception. With our logo printing service, you can add a personal touch to your kit, making it easily identifiable and enhancing your professional image.
Key Features:
Whether you're in need of a regular pressure transducer, a transducer with an integral flush device, or a disposable blood pressure transducer, our kit has you covered. It is compatible with patient monitors and is an essential accessory for accurate blood pressure measurement.
Don't compromise on quality when it comes to patient care. Choose our Disposable Pressure Transducer Kit for reliable and precise blood pressure monitoring. Order yours today and experience the difference!
Properties
|
Medical Materials & Accessories
|
Product Name
|
Abbott disposable IBP transducer
|
Brand Name
|
JCR
|
Model Number
|
JIBPT-01-YP
|
Main technical parameters and working environment requirements:
|
||||
Disposable Pressure Transducer
|
||||
Barometric pressure
|
70~106Kpa
|
|||
Relative humidity
|
10~90% (non-condensing)
|
|||
Operating pressure range
|
-50 ~ + 300 mmHg
|
|||
Sensitivity
|
5.0μV/V/mmHg ±3%
|
|||
Nonlinearity and hysteresis
|
full-scale reading±1.5%
|
|||
Input impedance
|
1200Ω ~3200Ω
|
|||
Zero pressure offset
|
-20mmHg~+20mmHg
|
|||
Thermal offset shift
|
≤±0.3mmHg/°C
|
|||
Offset drift
|
After warm-up for 20 seconds, drift within 2 mmHg within 8 hours
|
|||
Thermal span shift
|
≤±0.1%/°C
|
|||
Frequency response
|
Standard pressure set (48 "/ 12") is 40 Hz; Separate sensor > 200 Hz
|
|||
Defibrillator withstand
|
Defibrillation depends on the final connection of the equipment.
|
|||
Leakage current
|
Leakage current is dependent on the final connection of the equipment
|
|||
Overpressure load
|
-400~+4000mmHg
|
|||
Shock resistance
|
Withstand falling from one meter for three times
|
|||
|
When exposed to a 3400°K tungsten light, candle light at 3000 feet, it's less than 1 mmHg under rated voltage.
|
|||
Storage and transport environment requirements
|
- 20 ° C~ + 60 ° C, < 90%(non-condensing), no corrosive gas and indoor ventilation, and to avoid high temperature and cold.
|
|||
Contact with the human body
|
≤168h
|
|||
Intra-system velocity
|
||||
Flow
|
2-5mL/h
|
|||
Intra-system refers to the velocity under 6.00 VDC and 25°C, unless there are additional instructions.
|
||||
Product shelf-life
|
3 years
|